PH-METER GLP-21

PH-METER GLP 21 PH-METER GLP 21 IBERTEST

LABORATORY PH-METER GLP 21

Standards: EN-196-2

MAIN FEATURES
  • ph, mV and Temperature measurement.
  • Interchangeable probes.
  • Automatic calibration in 2-3 points with re-calibration warnings.
  • Control over magnetic or bar stirrer, velocity and on/off.
  • Interface RS 232C for printer or PC and communication software (optional)
  • Calibration with Standard buffers or by means any type of reference solution.
  • Password to protect the measuring program.
  • Internal Clock which the equipment offers calendar, chronometer and clock functions.
  • Displaying of the time needed for each measuring. An increment of this time has a direct relation with the electrode decay.
  • LCD Display with backlight liquid crystal.
  • Languages: Spanish, French, English and Italian.
DOUBLE WORKING MODE

1FUNCTIONAL MODE

The GLP-21 works as a classig pH-meter:
  • Ph measuring with resolution of 0.01 pH.
  • Reading by stability.
  • Calibration with two or three standards.
  • Data output to PC or Printer.

2ADVANCED MODE

The user can configure his own program, selecting:
  • Measuring units: pH or mV.
  • Measuring mode, by stability, in continuous operation, or by time.
  • Calibration with two or three buffers.
  • Buffers type, technical values or NIST values.
  • Measuring limits establishment, which facilitates a quick quality control of samples or some process displaying. An accoustic signal warns to the user about the measures out of preset limits.
  • Definition of one final point, with warning once reached.
TECHNICAL SPECIFICATIONS
Measuring Scales: pH -2 … + 16
mV ± 2000
ºC -20.0 … 150.0
Standard Resolution: 0,01 pH (Selectable 0,1 pH)
1 mV
0,1 ºC
Measuring error (± 1 digit) ≤ 0,02 pH
≤ 1 mV
≤ 0,3 ºC
Repeatability (± 1 digit) ± 0,01 pH, ± 1 mV, ± 0,1 ºC
Temperatura readjustment ± 0,01 pH, ± 1 mV, ± 0,1 ºC
Calibration Calibration in 2 or 3 points, validity between 0h and 7 days. Special calibration in any point between 0 and 14 pH. Recalibration warning. Forced calibration option. Automatic rejection of electrodes in poor condition.
Buffer identification Technicals: In 25 ºC: pH 2.00, 4.01, 7.00, 9.21, 10.90.
NIST in 25 ºC: pH: 1.679, 4.006, 6.865, 9.180,10.254
Screen display Alpha-numeric mode, LCD, backlight liquid crystal with 4 lines of 20 characters
Languages Spanish, Italian, Engish and French
Inputs and outputs Measuring electrode(BNC connector)
Reference electrode (banana connector Ø4)
Interface Standard PC keyboard
Interface RS 232 C, single way
Polarizing current, 10 μA ( as example, for Karl Fischer determination)
Analogical output, electrode potential follower
Control for the stirrer (Optional accesory): stop/run/velocity
Power supply External plug-in power supply, 12 VCC / 175 mA
Electrical security directives According to UNE-EN 61012
EMC According to CE, UNE-EN 50081-1 and UNE-EN 50082-1
Allowed environment conditions Working temperatura: 5 … 40 ºC.
Storing and transport temperature: -15 … 65ºC.
Relative Humidity < 80% not condensed
Container Made of high density polyurethane and enamelled steel.
Approx. weight 1.580 gr.
Dimmensions 180 x 215 x 130 mm