ELECTRONIC MOISTURE BALANCE EUROTHERM
Standards: EN-196-2
FUNCTIONS
- Max Sample Weight : 200 g.
- Not need of an exact sample amount.
- Display for Moisture Content (%) or Weight (g)
- Displays for Time (min) and Temperature (°C)
- All parameters programmed by keyboard.
- Time of tests programmable by 1 min steps.
- Furnace temperature programmable by 1 degree steps.
- Memory facility to retain Temperature, Time, Weight, Dry Residue and Moisture data until cancelled by the operator.
- Automatic Stable Weight determination and data storing at the end of the test.
- Software for memorisation of 5 cold weighings.
- Automatic stop at the end of the test with buzzer.
- Automatic calibration with external mass (optional)
SPECIFICATIONS
- Capacity: 200 g.
- Readability: 1 mg.
- Linearity: ± 1 mg.
- Repeatability: ± 0,5 mg.
- Sample weight: min 140 mg - max 200 g.
- Operating Temperature: 10°÷ 40°C.
- Timer Range: 1 ÷ 999 minutes.
- Furnace Temp. Range: 50°÷ 180°C
- Moisture accuracy: ± 0.01 % (sample min. 10 g)
- Moisture resolution: 0.01 % (sample min. 10 g)
- Data output: RS232 I/O adjustable.
- Pan diameter: 120 mm.
- Power requirements: 220 VAC (-15%/+10%) 50 Hz - 2A - 350 Watt.
- Dimensions (W x L x H): 210 x 355 x 300 mm.
- Net weight: 9 Kg.
STANDARD EQUIPMENT
- 3 calibrated and interchangeable stainless steel pans.
- 100 aluminium foil sheets.
- 1 pair of forceps.
ON REQUEST
- Calibrated mass E2 or F1 class (with or without certificate) from 50 g to multiples up to a maximum 200 g.
- Device for weighing underneath the moisture balance to determine humidity < 1%
- Software for data and statistics memorisation.